Quick Reference: Blender Failure
- Connect backup oxygen source / portable O2 tank directly to oxygenator.
- Have
someone go get the backup blender. - Disconnect the air and oxygen lines from the
wall. - Attach the backup blender to the pump and attach
its lines to the wall. - Remove the gas tubing from the failing blender
and attach it to the backup.