Hemodilution Increases the Susceptibility of Red Blood Cells to Mechanical Shear Stress During In Vitro Hemolysis Testing
Hemodilution with BSA maintains protein concentration within a physiologic range and reduces bias during hemolysis testing at high blood dilutions. Thus, American Society for Testing and Materials standards could consider including BSA as a diluent, when and as required: where large dilution is required (<83%) use PBS + 4 g% BSA, otherwise use PBS alone.